Additional information
Features | Conforms to ASTM D1533, D4928, D6304 Procedures A & B, IP 386, IP 438, API MPMS, Chap. 10.9, BS 60814, ISO 10101-3, ISO 10337, ISO 12937 |
---|---|
Test Method | Determines low concentrations of water in a wide range of liquid, gas and powder samples. Used for assessing water content in petroleum and petrochemical products including oils, gasolines, solvents, and fluids as well as other products such as pharmaceuticals and cosmetics. |
Specification | Titration method: End point detection: AC polarization End point indication: Visual display/print out/acoustic beep Display: 40 character alphanumeric backlit LCD Measuring range (possible): 1μg – 100mg water Measuring range (typical): 1μg – 10mg water Moisture range: 1 ppm – 100% Max. sensitivity: 0.1 μg Max. titration speed: 2.0 mg per minute Max. current: 400 ma Drift compensation: Automatically controlled Start delay time: 0 -30 minutes, user selectable End delay time: 0 – 30 minutes, user selectable Power supply: 90-264VAC, 47-63Hz Universal input Precision: 10-100μg ±3μg, 100μg-1mg ±5μg, above 1mg ±0.5% Calculation modes: Weight/weight, user programmable Display format: μg, mg/kg, ppm, % Print format: μg, mg/kg, ppm, % Statistics: max, mean, min values up to 99 runs Method storage: 10 user programmable methods Sample ID number: user programmable Printer: 42 character high-speed thermal printer Stirrer speed: Microprocessor controlled Language: English, Francais, Espanol, Portugues, Deutsch and Magyar Calendar/clock: Analysis time and date print out |
Electrical Requirements | 115-240V 50/60Hz |
Additional Accessories | K90850-1 Carry Case |
Dimensions | Dimensions: 250 x 245 x 120 mm |