Autoignition Apparatus

Autoignition Apparatus

Modified crucible furnace with digital thermocouple readout of flask temperature at prescribed points per ASTM specifications. Furnace provides rapid response and ±1°C stability throughout the operating range from ambient to 1200°C. The cylindrical heating chamber provides excellent radial temperature uniformity. The furnace cover has ports for flask exterior thermocouples, and a borosilicate glass thermocouple tube is provided to assure correct positioning of the gas temperature thermocouple inside the test flask. Thermocouples plug directly into the furnace control unit for quick disconnection when removing the flask. A hinged holder in the cover facilitates handling of the test flask. An adjustable mirror permits safe viewing of the flask interior during testing. The touch screen control panel has temperature controls and digital thermocouple readout with all temperatures displayed on a single screen.

$6,000.00

Description

Modified crucible furnace with digital thermocouple readout of flask temperature at prescribed points per ASTM specifications. Furnace provides rapid response and ±1°C stability throughout the operating range from ambient to 1200°C. The cylindrical heating chamber provides excellent radial temperature uniformity. The furnace cover has ports for flask exterior thermocouples, and a borosilicate glass thermocouple tube is provided to assure correct positioning of the gas temperature thermocouple inside the test flask. Thermocouples plug directly into the furnace control unit for quick disconnection when removing the flask. A hinged holder in the cover facilitates handling of the test flask. An adjustable mirror permits safe viewing of the flask interior during testing. The touch screen control panel has temperature controls and digital thermocouple readout with all temperatures displayed on a single screen.

Additional information

Test Method

Determines the lowest temperature at which the vapors of a liquid or solid chemical sample will self-ignite under prescribed laboratory conditions. The temperatures at which ‘cool flame’ and ‘hot flame’ ignitions occur, as evidenced by sudden temperature increases in the sample flask, are measured and recorded, and the delay time between introduction of the sample and ignition is timed.

Features

Conforms to ASTM E659 specifications
Digital furnace temperature control
Digital flask temperature display

Included Accessories

Borosilicate Test Flask, 500mL
Thermocouples (4)

Specification

Conforms to the specifications of: ASTM E659

Temperature Range: ambient to 1200°C

Temperature Control: digital setpoint touch screen control panel accurate to within ±1°C

Flask Temperature Display: 0-1200°C, via touch screen control panel

Electrical Requirements

220-240V 50/60Hz, Single Phase, 7.7A

Dimensions

WxDxH: 37x17x25 in (94×43.2×63.5 cm)
Net Weight: 85 lbs (38.6 kg)

Shipping Information

Shipping Weight: 125 lbs (56.7 kg)

Dimensions: 48x48x32 in (122x122x81.3 cm)